CHALLENGE
Energy saving version
Challenge "E" series have
basic environmental performances and require simple utilities
connection.
They are intended for general applications when high
environmental performances are not required.
An ideal tool for tropical tests or pharmaceutical products
(stability tests), plastic, rubber, paper, semiconductor
(85/85).
CHALLENGE "E" technical characteristics
|
Features |
Model
|
|
CH250 E
|
CH340 E
|
CH600 E
|
CH1200 E
|
>CH1600 E |
|
Useful capacity (liters) |
224
|
336
|
557
|
1152
|
1535
|
|
Internal dim (mm) (WxDxH) |
600x535x700
|
600x800x700
|
850x733x895
|
1000x1130x1020
|
1000x1505x 1020
|
|
External dim (mm) (WxDxH) |
850x1460x1516
|
850x1725x1516
|
1100x1705x1911
|
1250x2160x2036
|
1250x2534x 2040
|
|
Temperature range |
-20°C/180°C
|
-20°C/180°C
|
-20°C/180°C
|
-20°C/180°C
|
-20°C/180°C
|
|
Precision over time (T) |
±1°C
|
±1°C
|
±1°C
|
±1°C
|
±1°C
|
|
Heating rate: (0/+100°C) |
1.7°C/min
|
1.4°C/min
|
1°C/min
|
1°C/min
|
0.7°C/min
|
|
Cooling rate: (+100/0°C) |
1.7°C/min
|
1.4°C/min
|
1°C/min
|
0.5°C/min
|
0.3°C/min
|
|
Dissipation T=0°C |
>500 W
|
500 W
|
500 W
|
<500 W
|
500 W
|
|
R.H. range |
10%... 95%
|
10%... 95%
|
10%... 95%
|
10%... 95%
|
10%... 95%
|
|
R.H. precision |
±1%…±4%
|
±1%…±4%
|
±1%…±4%
|
±1%…±4%
|
±1%…±4%
|
|
Electrical power (average) |
2 kW
|
2 kW
|
2 kW
|
2.2 kW
|
2.2 kW
|
|
Weight (Kg) |
460
|
520
|
720
|
960
|
1000
|
|
Noise dB(A) |
59
|
59
|
59
|
59
|
59
|
|
Voltage |
230 V +6% / -10% / 50Hz / 1 ph + G
|
|